Abstract

Upper extremity arthroscopic surgery is a highly technique-dependent procedure that requires the surgeon to assess difficult cartilage conditions and manage the risk of iatrogenic damage to nerves and vessels adjacent to the joint capsule in a confined joint space, and a device that can safely assist in this procedure has been in demand. In this study, we developed a small intra-articular ultrasound (AUS) probe for upper extremity joint surgery, evaluated its safety using underwater sound field measurement, and tested its visualization with a phantom in which nerves and blood vessels were embedded. Sound field measurement experiments confirmed the biological safety of the AUS probe's output, while confirming that sufficient output power level performance was obtained as an ultrasound measurement probe. In addition, images of blood vessels and nerves were reconstructed discriminatively using A-mode imaging of the agar phantom. This study provides proof-of-concept of the AUS probe in upper extremity surgery. Further studies are needed to obtain approval for use in future medical devices.