Abstract

An international intercomparison study was carried out to optimise the readout procedures of the dosimetry systems used in the corresponding national environmental dosimetry laboratories. The intercomparison was based on the joint cross-calibration in the laboratories involved and an exchange of dosemeters within the respective national environment monitoring programmes. The aims of the study were the intercomparison of field irradiations at three exposure sites in each of the three participating countries, the determination of the actual field fading correction factor and the contribution of (national and international) transit doses. Six dosimetry systems have been employed in five measurement cycles using thermoluminescent LiF:Mg,Ti, CaF 2 :Mn, and CaSO 4 :Dy, as well as photoluminescent flat glass dosemeters. Within the uncertainty of measurement the results indicate no seasonal influence on the natural background level at any location, as well as no improvement in results by using fading corrections.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.