Abstract

In this work the least-squares background correction (LSBC) and internal standardization (IS) techniques were combined to eliminate spectral and transport interferences in the determination of Pb in phosphoric acid by high-resolution continuum source atomic absorption spectrometry (HR-CS AAS). Blanks, samples and reference solutions [0.10–1.00 mg L − 1 Pb in 1% (v/v) HNO 3] were spiked with 4.00 mg L − 1 Co used as internal standard. For absorbance measurements at the wavelength integrated absorbance equivalent to 9 pixels, correlations between the ratio of absorbance of Pb to absorbance of Co and the analyte concentration were close to 0.9992. Relative standard deviations of measurements varied from 0.6 to 4% and 1 to 7% ( n = 12) without and with IS/LSBC techniques, respectively. Recoveries for Pb spikes were in the 96–104% and 76–180% range with and without IS/LSBC, respectively. The limit of detection improved with IS/LSBC techniques. Accuracy of the proposed method was checked for the determinations of Pb in commercial phosphoric acid samples and results obtained with IS were better than those without IS.