Abstract

To demonstrate the results of the ray tracing-type aberrometer in measuring spherical aberration (SA) in pseudophakic eyes with monofocal intraocular lens (IOL), aspheric monofocal IOL, or aspheric diffractive multifocal IOL. Total, corneal, and internal SA were measured using iTrace at a 6-mm pupil size in 27 eyes of 27 patients implanted with a monofocal spherical IOL (group 1: Natural, SN60AT), 30 eyes of 30 patients implanted with a monofocal aspheric IOL (group 2: IQ, SN60WF), and 30 eyes of 30 patients implanted with a multifocal aspheric IOL (group 3: ReSTOR, SN6AD1) at 3 months after cataract surgery. We compared the internal SAs of these IOLs in pupil sizes of 3, 4, 5, and 6 mm. There were no demographic statistically significant differences among the groups. The internal SA of group 1 had a positive value. The internal SA of group 2 was -0.175 ± 0.135 μm in 5-mm pupils and -0.227 ± 0.253 μm in 6-mm pupils. The internal SA of group 3 was -0.072 ± 0.128 μm in 5-mm pupils and -0.173 ± 0.231 μm in 6-mm pupils. Measuring internal SA with iTrace yields relatively accurate results in all types of IOLs with adequate pupil sizes.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.