Abstract

This paper describes an equilibrium cell model for intermittent counter-current extraction that is analytically solved for the first time for continuous sample injection between a pair of columns. The model is compared with practice for injections of a model mixture of compounds on a standard high-performance counter-current chromatography instrument giving good agreement for compound elution order and the times to maximum concentration for the eluted components. An improved design of end fittings for the counter-current chromatography bobbins is described which permits on-column switching of the mobile and stationary phases. This on-column switching successfully eliminates the displaced stationary phase seen in fractions when operating ICcE with standard flying leads and gives a 6% reduction in the retention time of compounds and improved resolution due to the elimination of the time delay required to pump the previous mobile phase from standard flying leads.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.