Abstract

Analyses of statistical variance were applied to evaluate the precision and practicality of a CD-based NACE assay for R-timolol after enantiomeric separation of R- and S-timolol. Data were collected in an interlaboratory study by 11 participating laboratories located in Europe and North America. General qualitative method performance was examined using suitability descriptors (i.e. resolution, selectivity, migration times and S/N), while precision was determined by quantification of variances in the determination of R-timolol at four different impurity levels in S-timolol maleate samples. The interlaboratory trials were designed in accordance with the ISO guideline 5725-2. This allowed estimating for each sample, the different variances, i.e. between-laboratory (s2(Laboratories)), between-day (s2(Days)) and between-replicate (s2(Replicates)). The variances of repeatability (s2r) and reproducibility (s2R) were then calculated. The estimated uncertainty, derived from the precision estimates, seems to be concentration-dependent above a given threshold. This example of R-timolol illustrates how a laboratory can evaluate uncertainty in general.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.