Abstract

The effects of probe beam residual power in interferometry are investigated. It is seen that residual power produces an asymmetry between the forward and backward beams that is scaled with f, i.e., the effect can be easily seen in very slow optical systems. Common methods to detect residual power are discussed, showing the effectiveness of lateral shear interferometry to carefully set the collimation of the probe beam. A device approaching the order of magnitude established by Rayleigh's quarter wavelength rule is demonstrated.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.