Abstract

We measured the interference of carbonyl compounds, drugs, and other substances in human serum on the determination of creatinine by the two-point, fixed-time kinetic modification of the Jaffé reaction as well as by four enzymatic methods. We added known concentrations of the interfering substances to a solution of creatinine in water. For bilirubin, we used both pooled normal sera with added bilirubin and icteric patient sera. The magnitude of interferences varies widely from method to method. Carbonyl compounds, dopamine, cephalosporines, and bilirubin interfere with the Jaffé reaction. Bilirubin, creatine, dopamine, ascorbic acid, and sarcosine interfere with the enzymatic methods. We conclude that the elimination of interferences in the determination of creatinine has still not been achieved.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.