Abstract

A new glucose meter was developed employing a novel disposable glucose sensor strip comprising a nicotinamide adenine dinucleotide-glucose dehydrogenase (NAD-GDH) and a mixture of Fe compounds as a mediator. An iron complex, 5-(2,5-di(thiophen-2-yl)-1H-pyrrol-1-yl)-1,10-phenanthroline iron(III) chloride (Fe-PhenTPy), was synthesized as a new mediator for the NAD-GDH system. Due to the high oxidation potential of the mediator, the detection potential was tuned to be more closely fitted toward the enzyme reaction potential, less than 400 mV (vs. Ag/AgCl), by mixing with an additional iron mediator. The impedance spectrometry for the enzyme sensor containing the mixed mediators showed an enhanced charge transfer property. In addition, a new cartridge-type glucose meter was manufactured using effective aligned-electrodes, which showed an enhanced response compared with conventional electrode alignment. The proposed glucose sensor resulted in a wide dynamic range in the concentration range of 30 - 500 mg dL(-1) with a reduced interference effect and a good sensitivity of 0.57 μA mM(-1).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.