Abstract

An intercomparison of eye lens dosemeters has been conducted in terms of the quantity Hp(3). For the first time, besides photon radiation also beta radiation qualities were included. Three dosemeter types designed for the quantity Hp(3) and ten for Hp(0.07) took part in the intercomparison. As shown in a previous intercomparison for photon radiation only, the dosemeters designed for Hp(0.07) and calibrated in terms of Hp(3) performed well in photon radiation fields. But for beta radiation, it turned out that Hp(0.07) dosemeters over-responded up to a factor of 5 000 (with respect to the true Hp(3) dose) in the medium beta energy range (85Kr with a beta endpoint energy of 0.69 MeV), while some Hp(3) dosemeters performed quite well. For medium (57 keV) and high (662 keV) energy photon radiation, all dosemeter types showed response values well within the trumpet curve according to the current draft of ISO 14146.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.