Abstract
In this study we assess the interchangeability and statistical agreement of two prevalent instruments from the non-invasive “sniffer” method and compare their precision. Furthermore, we develop and validate an effective algorithm for aligning time series data from multiple instruments to remove the effects of variable and fixed time shifts from the instrument comparison. The CH4 and CO2 gas concentrations for both instruments were found to differ for population means (P<0.05) and intra-cow variation (precision) (P<0.05) and for inter-cow variation (P<0.05). The CH4 and CO2 gas concentrations from both instruments can be used interchangeably to increase statistical power for example, in genetic evaluations, provided sources of disagreement are corrected through calibration and standardisation. Additionally, averaging readings of cows over a longer period of time (one week) is an effective noise reduction technique which provides phenotypes with considerable inter-cow variation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.