Abstract
An Ion-Interaction RP-IIR-HPLC method was developed, able to simultaneously separate neutral, basic, and acidic pesticides in a single run. The proposed method was validated by comparison of the results obtained for the same sample by conventional RP-HPLC and GC/MS methods. For this purpose parametric t-, F-, t-paired, and t- for multiple samples and non parametric Wilcoxon matched-pair signed-rank tests were employed. The accuracy of the proposed IIR-RP-HPLC method was evaluated with respect to a lab-made reference sample of spiked tap water.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Liquid Chromatography & Related Technologies
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.