Abstract

The coupling between liquid chromatography and mass spectrometry with an APCI or ESI interface (in positive or negative mode) is used here for multi-residue analyses in natural waters, covering basic and neutral pesticides as well as acid pesticides. The methods developed are applied to drinking and, river waters after the samples are concentrated by liquid-liquid extraction or solid phase extraction on C18 cartridges. Comparisons are made between UV detection and mass spectrometry and between two chromatographic methods for acid substances. The quantitation limits range from 0.01 to 0.1 μg/l according to the substance.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.