Abstract
Relative intensity measurements in (υ′ = const, υ″) band progressions of the (B1Σu+ → X1Σg+) Lyman systems of H2 and HD were used to determine the relative sensitivity of a monochromator–detector combination in the wavelength region from 1100 to 1650 A. Additional measurements in three band progressions of the Lyman–Birge–Hopfield system of nitrogen yielded results that are in good agreement with the hydrogen data. The calibration range could then be extended to 1820 A.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.