Abstract
Integration of microfluidics with integrated circuit chip is a promising method to achieve lab-on-a-chip for its high-level miniaturization and portability in application. In this paper, a novel approach for integration of complementary metal–oxide–semiconductor image sensor and microwell array using the 3-D wafer-level chip scale package (3-D WLCSP) technology for biodetector is presented. The key process including silicon microwell array fabricating, wafer-level bonding, backside grinding, through-silicon via etching, redistribution layer, and ball grid array formation are developed. A high-density serpentine electrical circuit is integrated on the backside of the package to work as a temperature controller. After the fabrication process optimization, a yield rate of 94.6% is obtained. The reliability of the packages is characterized by thermal cycling and highly accelerated stress test temperature storage tests, and the results show a good package level reliability. The proposed 3-D WLCSP provides a low-cost and reliable solution for the fluidic biodetector integration.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IEEE Transactions on Components, Packaging and Manufacturing Technology
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.