Abstract

Accurate measurement of water vapor isotopes (δ18O and δ2H) is fundamental for advancing our understanding of the hydrological cycle and improving hydrological model accuracy. This study introduces an innovative calibration methodology using a controlled evaporation mixer (CEM) for determining stable isotopic ratios in atmospheric water vapor via cavity ring-down spectroscopy. The CEM technique reliably produces a stable water vapor stream, crucial for enhancing the precision and accuracy of isotopic measurements. Its rapid adaptation to changes in water vapor concentration and compatibility with different water standards enhance calibration reliability. Demonstrated reproducibility in generating water vapor across a broad concentration range from 900 to over 25,000 ppmv, coupled with a substantial reduction in memory effects, makes this approach highly effective in both laboratory and field settings. This calibration advancement greatly enhances research capabilities for continuous atmospheric water vapor analysis, providing deeper insights into hydrological processes and atmospheric dynamics.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.