Abstract

Objective Adults typically receive only one cochlear implant (CI) due to cost constraints, with a contralateral hearing aid recommended when there is aidable hearing. Standard hearing aids differ from a CI in terms of processing strategy and function as a separate entity, requiring the user to integrate the disparate signals. Integrated bimodal technology has recently been introduced to address this challenge. The aim of the study was to investigate the performance of unilateral CI users with and without an integrated bimodal fitting and determine whether binaural streaming technology offers additional benefit. Study sample Twenty-six CI users using integrated bimodal technology. Design Repeated measures where outcomes and user experience were assessed using a functional test battery more representative of real life listening (speech perception in noise tests, localisation test, tracking test) and the speech, spatial and qualities-of-hearing scale (SSQ). Results Bimodal outcomes were significantly better than for CI alone. Speech perception in noise improvements ranged from 1.4 dB to 3.5 dB depending on the location of speech and noise. The localisation and tracking tests, and the SSQ also showed significant improvements. Binaural streaming offered additional improvement (1.2 dB to 6.1 dB on the different speech tests). Conclusions Integrated bimodal and binaural streaming technology improved the performance of unilateral CI users.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call