Abstract

A square wave voltammetric measurement system with an 8 × 8 array of sensing pixels is integrated within a single chip for multi point and high-speed electrochemical analysis. The sensor was designed to operate at 400 ms/frame under the condition of 20 mV and 500 Hz of step increment and square wave frequency, respectively. Peak currents from the 8 × 8 pixel array of potassium ferricyanide were obtained at the potential range from −0.5 to 0.5 V. We also obtained a small variation of the measured concentration result of all 64 pixels under the concentration range from 0.6 to 6 mM of potassium ferricyanide standard solution.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.