Abstract

Real-time mass spectrometry (MS) has attracted increasing interest in environmental analysis due to its advantages in high time resolution, minimization of sampling artifact, and avoidance of time-consuming sample pretreatment. Among real-time MS methods, secondary electrospray ionization MS (SESI-MS) is showing great promise for the detection of organic compounds in atmospheric particulate matter. In this study, we demonstrated the working principles of secondary nanoelectrospray ionization (Sec-nESI) for real-time measurement of laboratory-generated organic aerosols using l-tartaric acid (TA) as a model compound. Factors affecting the detection of TA particles using a homemade Sec-nESI source coupled with a high-resolution mass spectrometer are systematically investigated. Temperature of ion transport capillary (ITC) was found to be the key factor in determining the ion signal intensity, which shows an increase of intensity by a factor of 100 from ITC temperature of 100–300 °C and could be attributed to more efficient desolvation and ionization. The characteristic fragment ion at m/z 72.99 was selected for quantitative analysis of TA at normalized collision energy of 50%, the optimal value applied during MS/MS analysis. Detection limit of 0.14 µg/m3 and a linear range of 0.2–2.97 µg/m3 are achieved. Satisfactory correlations between ion signal intensity and particle surface area (R2 = 0.969) and mass concentration (R2 = 0.967) were obtained. Although an equally good correlation was observed between signal intensity and particle surface area, the good correlation between signal intensity and particle mass concentration indicates that high solubility of TA ensures efficient dissolution of TA in the primary ESI droplets for further ionization.

Highlights

  • Recent developments in real-time chemical characterization of aerosol particles have garnered attention for their high time resolutions (Gallimore and Kalberer 2013; Li et al 2015, 2017b; Brüggemann et al 2015)

  • We investigated the real-time measurement of laboratorygenerated tartaric acid (TA) particles using a custom-built Sec-nESI source coupled with a high-resolution Quadrupole Orbitrap Mass Spectrometer (HRMS)

  • Parameters involved in the Sec-nESI process, including ESI voltage, ESI solution flow rate, sample flow rate, and ion transport capillary (ITC) temperature, were optimized

Read more

Summary

Introduction

Recent developments in real-time chemical characterization of aerosol particles have garnered attention for their high time resolutions (seconds to minutes) (Gallimore and Kalberer 2013; Li et al 2015, 2017b; Brüggemann et al 2015). Compared to off-line methods, these real-time techniques can minimize artifacts due to condensation and evaporation, and avoid time-consuming sample pretreatment (Brüggemann et al 2015). Advantages such as fast data acquisition and high sensitivity have made mass spectrometry (MS) among the most popular techniques for real-time aerosol chemical characterization (Pratt and Prather 2012; Li et al 2015, 2017b). IC-based methods are designed for water-soluble components (e.g., inorganic ions and very small carboxylic acids) and are not suitable for many hydrophobic organic compounds present in atmospheric particles (Li et al 2017b)

Methods
Results
Conclusion
Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call