Abstract

A detailed comparative analysis of several possible input circuits for observing NMR free precession signals is given. Special attention is paid to the case of signals having short decay time constants. Here the presence of a large initial ringing transient makes it necessary to consider possible modifications in the design of the receiver system. The analysis indicates that an untuned transformer coupled input circuit is potentially superior to the traditional parallel resonant circuit in this case.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.