Abstract

To assess image quality of virtual monochromatic spectral (VMS) images, compared to single-energy (SE) CT, and to evaluate the feasibility of material density imaging in abdominal aortic disease. In this retrospective study, single-source (ss) dual-energy (DE) CT of the aorto-iliac system in 35 patients (32 male, mean age 76.5 years) was compared to SE-CT. By post-processing the data from ssDECT, VMS images at different energies and material density water (WD) images were generated. The image quality parameters were rated on 5-point scales. The aorto-iliac attenuation and contrast-to-noise ratio (CNR) were recorded. Quality of WD images was compared to true unenhanced (TNE) images. Radiation dose was recorded and statistical analysis was performed. Image quality and noise were better at 70 keV (P < 0.01). Renal artery branch visualisation was better at 50 keV (P < 0.005). Attenuation and CNR were higher at 50 and 70 keV (P < 0.0001). The WD images had diagnostic quality but higher noise than TNE images (P < 0.0001). Radiation dose was lower using single-phase ssDECT compared to dual-phase SE-CT (P < 0.0001). 70-keV images from ssDECT provide higher contrast enhancement and improved image quality for aorto-iliac CT when compared to SE-CT at 120 kVp. WD images are an effective substitute for TNE images with a potential for dose reduction.