Abstract

Although emerging data support the utility of real-time three-dimensional echocardiography (RT3DE) during dobutamine stress testing, the feasibility of performing contrast enhanced RT3DE during exercise treadmill stress has not been explored. Two-dimensional (2D) and three-dimensional (3D) acquisition were performed in 39 patients at rest and peak exercise. Contrast was used in 29 patients (74%). Reconstruction was performed manually by generating short axis cut planes at the base, mid-ventricle and apex, and automatically by generating 9 short axis slices. Three-dimensional acquisition was feasible during rest and stress regardless of the use of contrast. Time to acquire stress images was reduced using 3D (35.2±17.9 s) as compared to 2D acquisition (51.6±14.7 s; P<0.05). Using a 17-segment model, of all 663 segments, 588 resting (88.6%) and 563 stress segments (84.9%) were adequately visualized using manually reconstructed 3D data, compared with 618 resting (93.2%) and 606 stress segments (91.4%) using 2D data (P rest=0.06; P stress=0.07). We concluded that contrast enhanced RT3DE is feasible during treadmill stress echocardiography.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.