Abstract
The focus of this research was to calculate initial beam characteristics of a 6 MV X-rays with flattening filter (FF) from Elekta Versa-HD linear accelerator (Linac) using the Tool for Particle Simulation (TOPAS) Monte Carlo (MC) simulation code. TOPAS 3.6 simulated data was compared with the measured data. During simulation, all the main component were simulated including Linac head, Multi-Leaf Collimators (MLCs), and uniform water phantom, while the SNC 3D scanner phantom and SNC125c Chamber were used for measurements. At collimator sizes of 5x5, 10x10, and 20x20 cm2, the depth dose scans (PDD) distribution and beam profiles (BP) were measured. The estimated tissue phantom ratio (TPR20,10) was 0.669, which was -1.1 percent lower than the measured TPR20,10 of 0.681. The obtained results show that the simulated and measured data were in good agreement, with gamma rate analysis of 3%/3 mm criteria reaching 95% and 99 % for PDD and BP respectively.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.