Abstract

We have investigated the effects of X-ray radiation of doses of 0, 200 Gy, and 20 kGy on the Hamamatsu silicon-photomultiplier (SiPM) S10362-11-050P. From current-voltage, capacitance/conductance-voltage, pulse-shape and pulse-height measurements, the SiPM characteristics below and above breakdown voltage were determined. Significant changes are observed, however, the changes above the breakdown voltage hardly affect the operation of SiPMs. We conclude that the operation of the Hamamatsu SiPM as high-gain photodetector is hardly affected by X-ray radiation damage in the dose range investigated.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.