Abstract
The 239Pu fission chambers are widely used to measure fission spectrum neutron flux due to a flat response to fast neutrons. However, in the meantime the resonance and thermal neutrons can cause a significant influence on the measurement if they are moderated, which could be eliminated by using 10B and Cd covers. At a column enriched uranium fast neutron critical assembly, the fission reaction rates of 239Pu are measured as 1.791×10−16, 2.350×10−16 and 1.385×10−15 per second for 15 mm thick 10B cover, 0.5 mm thick Cd cover, and no cover respectively, while the fission reaction rate of 239Pu is rapidly increased to 2.569×10−14 for a 20 mm thick polythene covering fission chamber. The average 239Pu fission cross-section of thermal and resonance neutrons is calculated to be 500 b and 24.95 b with the assumption of 1/v and 1/E spectra respectively, then thermal, resonance and fast neutron flux are achieved to be 2.30×106, 2.24×106 and 1.04×108 cm−2·s−1.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.