Abstract
The influence of radiation damage on energy resolution, linearity and uniformity of an electromagnetic lead/scintillator calorimeter was studied experimentally with 1–6 GeV electrons. Plastic scintillators and wavelength shifter bars were irradiated uniformly with γ rays. Both were identical with those of the ZEUS uranium calorimeter. The attenuation length of the scintillators was determined from bench tests of single scintillator tiles and from beam tests of the whole calorimeter. After exposure to a dose of 10 kGy the attenuation length of the scintillators decreased by a factor of 2. The experimental results show that the irradiation of the nearly 2 m long wavelength shifter bars affects the calorimeter much more than the damage of the scintillators which were only 19 cm long. Damaged and undamaged sections of the calorimeter were scanned with a moving radioactive γ source ( 60Co). The results demonstrate that the 60Co monitor system is a very precise tool to detect radiation damage in a sampling calorimeter.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.