Abstract

The objective of this study was to investigate the effects of eccentric cyclic loading on implant components using the internal joint system with titanium and zirconia abutments. Abutments were made of either pure titanium (Ti group) or zirconia (TZP group). Cyclic loading test was conducted according to the specifications of ISO 14801. Loading condition was at 2 points assuming axial load and eccentric load. The reverse torque value reduced after the eccentric load and reduced more in the TZP group than the Ti group. Marginal gap changed after eccentric loading, and was greater in the TZP group. In the TZP group, changes in configuration were observed in the implant body, and Ti was detected on the abutment surface. From the above, the eccentric load may have worse effects than axial loads. It was suggested that the TZP group was clinically disadvantageous compared to the Ti group.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.