Abstract

To verify the influence of the software version of the dose control system on the equivalent dose and effective dose calculation values. We performed chest CT and liver to pelvic CT imaging with a human phantom placed on a CT bed. After the imaging was completed, the radiation dose structured reports (RDSRs) generated by the equipment were transferred to several dose management systems with different software versions for equivalent dose and effective dose calculations. The equivalent and effective doses calculated from the same RDSR differed depending on the software version of the dose management system with a difference of approximately 2 times for the effective dose and up to 50 times for the equivalent dose. It is considered that the voxel phantom geometry and dose calculation algorithm may differ depending on the software version of the dose management system. Careful attention should be paid in handling the dose calculation values because the exposure explanations and risk assessments using the equivalent doses, and effective dose calculated by the dose management systems may be overestimated.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.