Abstract

The behavior of a CoCrMo alloy and its components was studied in simulated physiological solution (Hank’s solution) using cyclic voltammetry (CV) and electrochemical impedance spectroscopy (EIS). The passivation of all samples occurred spontaneously at the open circuit potential. The composition of the oxide films as a function of the solution components and the applied potential is discussed. The electrochemical properties of the spontaneously passivated electrodes at the open circuit potential were studied by EIS. The polarization resistance ( R p) and the electrode capacitance ( C dl) were determined. The higher R p of the alloy than of the chromium pointed to the stabilizing effect of the other alloy components. The concentration of the metallic ions in a simulated physiological solution, measured by inductively coupled plasma-mass spectrometry, was in accordance with the values of both the R p determined from EIS data and current densities measured with CV.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.