The ease of disinfection of the sample ports of three types of urine drainage leg bags with different sampling port systems was assessed using a bladder bag model. The ports were contaminated with Escherichia coli, ‘disinfected’ using a standard method, then sampled at time intervals up to one week after contamination. It was discovered that leg bags which employ a needle-based sampling system (the ‘sample safe port system’) were easier to disinfect than those which did not, and that organisms are retained in large enough numbers to lead to misdiagnosis of a urinary tract infection or to pose a retrograde infection risk.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.