Abstract
BackgroundRadiotherapy has been shown to cause malfunction of implantable cardioverter-defibrillators, and there are few studies of implantable cardioverter-defibrillators and radiotherapy. We report an unusual case of small cell lung cancer in a patient with an implantable cardioverter-defibrillator in whom direct irradiation to the electrode and lead could not be avoided.Case presentationWe report a case of radiotherapy in a 72-year-old Korean man with a limited stage of small cell lung cancer who had undergone insertion of an implantable cardioverter-defibrillator because of ventricular fibrillation. The radiation dose was 60 Gy in 30 fractions to the thorax. The mean dose and maximum dose estimated at the body of the implantable cardioverter-defibrillator were 0.89 Gy and 2.23 Gy, respectively. The mean and maximum doses of the lead and electrode were 17.12 Gy and 55.72 Gy in the lead and 1.81 Gy and 7.10 Gy in the electrode, respectively, because part of the lead and electrode was inevitably in the irradiated fields. The function of the patient’s implantable cardioverter-defibrillator was checked daily, and no change in implantable cardioverter-defibrillator function was observed for the duration of radiotherapy. The patient was tolerated the treatment well without severe complications. Computed tomography performed at 4 weeks after radiotherapy showed a good response with regression of the tumor. The patient was alive with complete remission of the tumor and without any implantable cardioverter-defibrillator dysfunction more than 36 months after the end of treatment.ConclusionsThis case demonstrates that radiotherapy may be a safe and effective treatment modality through careful monitoring of implantable cardioverter-defibrillators in patients with lung cancer who have implantable cardioverter-defibrillators.
Highlights
Radiotherapy has been shown to cause malfunction of implantable cardioverter-defibrillators, and there are few studies of implantable cardioverter-defibrillators and radiotherapy
This case demonstrates that radiotherapy may be a safe and effective treatment modality through careful monitoring of implantable cardioverter-defibrillators in patients with lung cancer who have implantable cardioverter-defibrillators
As the use of cardiac implantable electronic devices (CIEDs) such as permanent pacemakers or implantable cardioverter-defibrillators (ICDs) in the management of cardiovascular disease has increased with increasing life expectancy, so has the indication of radiotherapy in comorbidity of cancer and cardiovascular disease with Cardiac implantable electronic device (CIED) [1]
Summary
Radiotherapy has been shown to cause malfunction of implantable cardioverter-defibrillators, and there are few studies of implantable cardioverter-defibrillators and radiotherapy. Conclusions: This case demonstrates that radiotherapy may be a safe and effective treatment modality through careful monitoring of implantable cardioverter-defibrillators in patients with lung cancer who have implantable cardioverter-defibrillators. ICDs are composed of a body (generator) and wires (electrode and lead), these reports were focused on the pacemaker or the body of the ICD.
Published Version (
Free)
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have