Abstract

The induction of pulmonary hemorrhage by pulsed ultrasound was discovered over 20 years ago. This phenomenon may pose a risk of patient lung injury, particularly for point of care pulmonary ultrasound. A diagnostic ultrasound machine (HDI 5000; Philips Healthcare, Andover MA USA) with 7.6 MHz (CL15-7) linear array was used to image the right lung of anesthetized rats in a warmed water bath. The image showed rapid initiation and progression of comet tail artifacts across the lung image for an on-screen mechanical index (MI) of 0.9, which corresponded to a pulmonary hemorrhage in the lung. Groups of rats were scanned at a range of MI settings and a threshold was located at an MI of about 0.44. This finding indicated a greater sensitivity to pulmonary ultrasound than was expected, based on previous results. Further research is needed to understand this phenomenon and to develop safety guidelines for sonographers.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.