Abstract

Flow injection methods for the individual and simultaneous determination of ascorbic acid and uric acid are proposed. A spectrophotometer and a miniamperometric detector are connected in sequence. The calibration graphs for uric acid obtained by measuring its absorbance at 293 nm and its current at +0.6 V are linear up to at least 80 and 70 μg/ml, respectively, with an rsd ( n = 10) of 1 % for both methods at mid-range concentrations. The calibration graph for ascorbic acid with amperometric detection is linear up to 80 mg/l. with an rsd ( n = 10) of 0.8% at 30 mg/l. The simultaneous determination of uric acid and ascorbic acid is based on measurement of the absorbance of uric acid at 393 nm and amperometric determination of both analytes at +0.6 V. The average relative errors of the analysis of binary mixtures of uric acid and ascorbic acid are 2.2 and 4.2%, respectively.