Abstract

A rapid, indirect gas-diffusion flow injection analysis (FIA) method with amperometric detection has been developed for the selective and sensitive determination of tetrahydroborate (BH − 4). The injected analyte reduces arsenic(III) to arsine. The arsine formed diffuses through the PTFE (polytetrafluoroethylene) membrane and is quantified amperometrically at a platinum working electrode. The precision of the technique was better than a relative standard deviation of 2.1% at 60 μ M levels and better than 0.5% at 0.1 m M, with a throughput of 60 samples/hr. The detection limit of the method was found to be 1 μ M (1.5 ng BH − 4) with a linear range up to 1 m M. The dynamic range extends over five orders of magnitude in BH − 4 concentration. The effects of working potential, concentration of As(III) and HCl in the reagent stream, type and flow rate of the acceptor solution, temperature and interferents on the FIA signals were studied.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.