Abstract

The application of an automated flow injection analysis (FIA) system for on-line determination of microbial activity, during high cell density cultivations of Escherichia coli is reported. Based on a bioelectrochemical principle, the FIA method used a redox mediator (potassium hexacyanoferrate(III)) to facilitate electron transfer from the microorganisms to an electrochemical detector. Assays were carried out using a new sampling device which provided aseptic operation by use of a valve and chemical sterilisation. No sample dilution or pretreatment was necessary for biomass concentrations up to approx. 40 g l −1. The sample volume was 0.5 ml and the overall analysis time was 5 min. FIA signals were found to correlate well with the oxygen uptake rate (OUR). Changes in metabolic activity due to low substrate levels or high inhibitor concentrations in the cultivation medium became obvious from the FIA signals.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.