Abstract

We investigated the basic characteristics of indirect arterial blood pressure (ABP) measurement using a device that combines oscillometry and photoplethysmography in cats. Dobutamine was infused intravenously in four anesthetized cats. Direct ABP was measured by a catheter. Indirect ABP was measured from the left forelimb. Dobutamine significantly elevated both systolic arterial pressure (SAP) and mean arterial pressure (MAP) in a dose-dependent manner. The indirect SAP, MAP and diastolic arterial pressure (DAP) values were closely correlated with the direct ABP values (r=0.88, 0.89 and 0.83, respectively). The mean bias for SAP, MAP and DAP was 3.4, 0.2 and −2.4 mmHg, respectively. The indirect ABP measured by this device may be used to reliably monitor ABP changes in anesthetized cats.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.