Abstract

To determine the level of agreement between direct and indirect blood pressure measurements obtained from healthy Hispaniolan Amazon parrots (Amazona ventralis) anesthetized with isoflurane. Validation study. 16 healthy adult Hispaniolan Amazon parrots. Parrots were anesthetized, and a 26-gauge, 19-mm catheter was placed percutaneously in the superficial ulnar artery for direct measurement of systolic, mean, and diastolic arterial pressures. Indirect blood pressure measurements were obtained with a Doppler ultrasonic flow detector and an oscillometric unit. The Bland-Altman method was used to compare direct and indirect blood pressure values. There was substantial disagreement between direct systolic arterial blood pressure and indirect blood pressure measurements obtained with the Doppler detector from the wing (bias, 24 mm Hg; limits of agreement, -37 to 85 mm Hg) and from the leg (bias, 14 mm Hg; limits of agreement, -14 to 42 mm Hg). Attempts to obtain indirect blood pressure measurements with the oscillometric unit were unsuccessful. Results suggested that there was substantial disagreement between indirect blood pressure measurements obtained with a Doppler ultrasonic flow detector in anesthetized Hispaniolan Amazon parrots and directly measured systolic arterial blood pressure.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.