Abstract

Deviations in ultrafiltration volume can cause various problems for patients undergoing a haemodialysis treatment. With the introduction of high flux dialysers, closed systems with volumetric fluid balancing became the systems of choice, because control of transmembrane pressure used previously is too inaccurate and unsafe to be used with such filters. In this article, a reliable and cost effective test method is described to increase the safety of a closed system. The test is carried out by simply analysing the dialysate pressure while the dialyser is isolated from the system for typically 7.2 seconds. No additional sensors and components other then a valve combination used to isolate the dialyser are required.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.