Abstract

Abstract In continuation of our work dealing with multicolumn HPLC (MC[sbnd]HPLC) we describe in this paper an on-line on-column fraction trapping technique based on effluent mixing. To a normal two-column switching set-up (in this case with two RP columns) an additional high-pressure pump gets inserted into the connection line between column A and column B via a low dead volume mixing tee. The in-line respectively off-line switching of pump B and the mobile phase B is time controlled by using a high pressure switching valve. With this set-up it is possible to mix on-line an effluent fraction from column A and transferred onto column B with a highly polar and pH-controlled (e.g. aqueous buffer) new effluent, to reduce or adjust significantly the overall elution strength of this mixed transferred solvent. Thus, several chromatographically effective possibilities can be created in a simple manner, which are for example: (a) pronounced peak compression respectively on-column concentration on column B; (b) du...