Abstract

Surface passivation is critical for effective PCR using silicon-glass chips. We tested a dynamic polymer-based surface passivation method. Polyethylene glycol 8000 (PEG 8000) or polyvinylpyrrolidone 40 (PVP-40) applied at 0.75% (w/v) in the reaction mixture produced significant surface passivation effects using either native or SiO2-precoated silicon-glass chips. PCR amplification was achieved from human genomic DNA as a template as well as from human lymphocytes. The dynamic surface passivation effect of PEG 8000 remained similar under both conditions. Dynamic surface passivation offers a simple and cost-effective method to make microfabricated silicon-glass chips PCR friendly. It can also be used in combination with static passivation (silicon oxide surface layer) to further improve PCR performance using silicon-glass PCR chips.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.