Abstract

Incomplete filter opening, causing filter migration and inferior vena cava (IVC) perforation, have often been observed with the Greenfield-type IVC filter, but reports of incomplete opening of the ALN-type IVC filter are extremely rare. We present herein the case of incomplete opening of an ALN-type IVC filter due to entanglement of the filter legs, which caused filter migration, IVC wall perforation, and penetration into the kidney and iliopsoas muscle. Successful percutaneous retrieval was performed without complications. Our experience provides guidance for similar situations in the future, including the need to check for complete filter opening.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.