Abstract

The Vario function is a feature of certain pacemakers that allows noninvasive voltage threshold determinations. 1 After programming the Vario option, magnet application over the pulse generator results in 16 asynchronous impulses at a frequency of Wmin (new unit), followed by 15 asynchronous impulses at a frequency of 120/min., A stepwise decrease in amplitude occurs during these latter 15 cycles, followed by a final 0 volt output signal. The decrement in voltage per cycle is determined by the programmed amplitude as well as battery status. In a new unit, at a 5 volt output, the step differential is 0.32 volts. Initiation of the Vario test results from engagement of a reed switch by the test magnet. A case is described where failure of the reed switch to disengage after magnet removal led to continuous Vario cycling.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.