Abstract

A perturbed gut microbiome has recently been linked with multiple disease processes, yet researchers currently lack tools that can provide in vivo, quantitative, and real-time insight into these processes and associated host-microbe interactions. We propose an in vivo wireless implant for monitoring gastrointestinal tract redox states using oxidation-reduction potentials (ORP). The implant is powered and conveniently interrogated via ultrasonic waves. We engineer the sensor electronics, electrodes, and encapsulation materials for robustness in vivo, and integrate them into an implant that endures autoclave sterilization and measures ORP for 12 days implanted in the cecum of a live rat. The presented implant platform paves the way for long-term experimental testing of biological hypotheses, offering new opportunities for understanding gut redox pathophysiology mechanisms, and facilitating translation to disease diagnosis and treatment applications.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.