Abstract

Purpose Iontophoresis kills microbes in vitro and, therefore, may be a useful method for eliminating microbial populations associated with catheter-induced urinary tract infections in vivo. Materials and Methods Catheters were modified to deliver current to platinum electrodes in the catheter tip. Female sheep were catheterized with this iontophoretic catheter and left ambulatory. In 5 sheep (experimental group) 400 microampere was applied to the catheter and withheld in 4 sheep (control group) for 20 to 21 days. The animals were then sacrificed. During the study, types and concentrations of bacteria, and physical and chemical characteristics of the urine samples were determined. Results Throughout the study, bacteria levels were reduced in urinary tracts of the experimental group (10 3 to 10 4 microbes per ml.) compared with the control group (10 7 microbes per ml.), without extensive alterations to urine chemistry or the sheep urinary tract. Conclusions Since iontophoresis safely reduced bacterial populations in catheterized sheep, this technology may reduce or eliminate nosocomial, catheter-induced urinary tract infections in humans.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call