Abstract

To evaluate the volume and heat-sink effects of microwave ablation (MWA) in the ablation zone of the normal swine lung. MWA at 100 W was performed for 1, 2, and 3 min in 7, 5, and 5 lung zones, respectively. We assessed the histopathology in the ablation zones and other outcome measures: namely, length of the longest long and short axes, sphericity, ellipsoid area, and ellipsoid volume. The mean long- and short-axis diameters were 22.0 and 14.1 mm in the 1-min ablation zone, 27.6 and 20.2 mm in the 2-min ablation zone; and 29.2 and 21.2 mm in the 3-min ablation zone, respectively. All measures, except sphericity, were significantly less with 1-min ablation than with either 2- or 3-min ablation. There were no significant differences between the 2- and 3-min ablation zones, but all measures except sphericity were larger with 3-min ablation. Although there were no blood vessels that resulted in a heat-sink effect within the ablation zones, the presence of bronchi nearby in 5 lung ablation zones resulted in reduced ablation size. In high-power, short-duration MWA, the lung ablation volume was affected by ablation time. Some ablations showed that a heat-sink effect by a neighboring bronchus might occur.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.