Abstract

The aim of this study was to compare the differences between the ablation region and hyperechoic zones in microwave and radio frequency ablation of different tissues. Microwave and radio frequency ablation were performed on fresh porcine muscle and liver with different power levels for 90 seconds. These 2 ablation methods were then performed on rabbit liver in vivo using 20 W for 60 seconds. The volumes of the ablation and hyperechoic zones were compared following different ablation methods. The ablation zones were significantly greater than the hyperechoic zones (P < .05) with the same power and duration when using 2 ablation methods. The differences of the ablation and hyperechoic zones between muscle and liver tissues were significantly different (P < .05). The difference values of the ablation and hyperechoic zones were also significantly different (P < .05) using 2 ablation methods. The hyperechoic zone may have underestimated the extent of ablation using a specified ablation time. In the same tissue, the hyperechoic zone could more accurately estimate the ablation zones using microwave ablation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.