Abstract
Ultrasound-based local pulse wave velocity (PWV) estimation, as a measure of arterial stiffness, can be based on fast focused imaging (FFI) or plane wave imaging (PWI). This study was aimed at comparing the accuracy of in vivo PWV estimation using FFI and PWI. Ultrasound radiofrequency data of carotid arteries were acquired in 14 healthy volunteers (25-57 y) by executing the FFI (12 lines, 7200 Hz) and PWI (128 lines, 2000 Hz) methods consecutively. PWV was derived at two time-reference points, dicrotic notch (DN) and systolic foot (SF), for multiple pressure cycles by fitting a linear function through the positions of the peaks of low-pass filtered wall acceleration curves as a function of time. The accuracy of PWV estimation was determined for various cutoff frequencies (10-200 Hz). No statistically significant difference was observed between PWVs estimated by both approaches. The PWV and R2 at DN were higher, on average, than those at SF (PWV/R2: FFI SF 5.5/0.92, FFI DN 6.1/0.92; PWI SF 5.4/0.89, PWI DN 6.3/0.95). The use of cutoff frequencies between 40 and 80Hz provided the most accurate PWVs. Both methods seemed equally suitable for use in clinical practice, although we have a preference for the PWV at DN given the higher R2 values.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.