Abstract
ObjectiveThis in vitro study was undertaken to evaluate the physical, chemical, and biological properties of commercially available metal orthodontic brackets in South Korea, because national standards for these products are lacking.MethodsFour bracket brands were tested for dimensional accuracy, (manufacturing errors in angulation and torque), cytotoxicity, composition, elution, and corrosion: Archist (Daeseung Medical), Victory (3M Unitek), Kosaka (Tomy), and Confidence (Shinye Odontology Materials).ResultsThe tested rackets showed no significant differences in manufacturing errors in angulation, but Confidence brackets showed a significant difference in manufacturing errors in torque. None of the brackets were cytotoxic to mouse fibroblasts. The metal ion components did not show a regular increasing or decreasing trend of elution over time, but the volume of the total eluted metal ions increased: Archist brackets had the maximal Cr elution and Confidence brackets appeared to have the largest volume of total eluted metal ions because of excessive Ni elution. Confidence brackets showed the lowest corrosion resistance during potentiodynamic polarization.ConclusionsThe results of this study could potentially be applied in establishing national standards for metal orthodontic brackets and in evaluating commercially available products.
Published Version (Free)
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have