Abstract
Abstract Total cervical disc replacement aims at avoiding fusion for the treatment of degenerative disc disease. This paper investigates the wear performances of PEEK and CFRPEEK against UHMWPE for artificial cervical disc application on a knee wear simulator due to the unavailability of a spinal one. The obtained results showed that the average wear rates of PEEK-on-UHMWPE and CFRPEEK-on-UHMWPE were 3.82 mg/MC and 2.34 mg/MC, respectively. The edge area of the bearing surfaces displayed severer wear in comparison with the central zone and was dominated by arc-shaped wear tracks. The main wear mechanisms were abrasion, adhesion and fatigue wear. CFRPEEK-on-UHMWPE has demonstrated superior wear characteristics, and hence, is a potential material configuration for artificial cervical disc application.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.