Abstract

This paper investigates the possibility of robotically performing in situ needle manipulations to correct the needle tip position in the setting of robot-assisted, MRI-guided spinal injections, where real time MRI images cannot be effectively used to guide the needle. Open-loop control of the needle tip is derived from finite element simulation, and the proposed method is tested with ex vivo animal muscle tissues and validated by cone beam computed tomography. Preliminary results have shown promise of performing needle tip correction in situ to improve needle insertion accuracy when real-time feedback is not readily available.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.